Summary
Type: SPE
Application #: 303820
Catalog#: 730004 - SPE Cartridges, Chromabond, C18, 6ml, PP, 500mg, 30/pk
Detection: HPLC in reference to Appl. No.: 121820 on NUCLEOSIL 100-5 NH2
Conditions: Column conditioning: 5 ml methanol, then 20 ml water.
(Do not let run the column dry.)
Sample aspiration: The prepared sample is passed through the column
Washing: 20 ml water, then 15 ml (30% methanol in water)
Elution: 20 ml methanol
Evaporation to dryness and reconstitution with 2 ml methanol
Authors: Productmanagement Chromatographie (2007) based on Ph Eur
Source: Macherey-Nagel